English Abstract This report deals with the performance characteristics
of a method by means of ICP-AES-axial technique (PE Optima 3000 XL). The
already existing method 'CuCrNi' was tested for the detection limit,
selectivity, repeatability, reproducibility, accuracy, linearity, drift
between successive calibrations and recovery. The results are shown for
eight spectral lines: three of chromium, two of copper and three of nickel.
In all measurements correction for matrix effects was carried out using an
internal standard (indium) and proportional correction. As most important
results are mentioned here the detection limits: 0,004, 0,004 and 0,007
mumol/l and the average recovery of an analyte addition: 103.5%, 102.0%
and 101.0% for Cr, Cu and Ni respectively.
Rapport in het kort
Dit rapport behandelt de prestatiekenmerken van een
methode met behulp van ICP-AES met een axiaal plasma (Perkin-Elmer Optima
3000 XL). Met de reeds bestaande methode 'CuCrNi' werden de volgende
prestatiekenmerken vastgesteld: aantoonbaarheidsgrens, selectiviteit,
herhaalbaarheid, reproduceerbaarheid, juistheid aan de hand van de analyse
van een referentiemateriaal, lineariteit van de kalibratielijn, verloop van
de kalibratielijn en terugvinding van een toevoeging van analiet aan
grondwatermonsters. Resultaten van acht analyselijnen worden gepresenteerd,
drie chroomlijnen, twee koperlijnen en drie nikkellijnen. Voor alle
metingen werd correctie voor matrixeffecten toegepast door het gebruik van
een interne standaard (indium) en door middel van proportionele correctie.
Als belangrijkste resultaten worden hier vermeld de aantoonbaarheidsgrenzen:
0,004, 0,004 en 0,007 mumol/l en de gemiddelde terugvinding van een
analiettoevoeging: 103,5%, 102,0% en 101,0% voor respectievelijk Cr, Cu en
Ni.