English Abstract This report deals with the performance characteristics
of a method for the determination of total phosphorus in groundwater by
means of ICP-AES-axial technique (PE Optima 3000 XL). The already existing
method 'TotP' was tested for the detection limit, repeatability,
reproducibility, linearity, selectivity, drift between successive
calibrations, recovery of phosphorus added to groundwater samples and of
organically bound phosphorus. The results of the analyte lines P178 and
P213, each with and without the use of the internal standard elements Ga and
In, were compared. The performance using the P213 analyte line in
combination with the internal standard Ga showed to be superior in terms of
sensitivity, detection limit and repeatability.
Rapport in het kort
Dit rapport behandelt de prestatiekenmerken van een
methode voor de bepaling van totaal-fosfor in grondwater met behulp van
ICP-AES met een axiaal plasma (Perkin-Elmer Optima 3000 XL). Met de reeds
bestaande methode 'TotP' werden de volgende prestatiekenmerken vastgesteld:
aantoonbaarheidsgrens, herhaalbaarheid, reproduceerbaarheid, lineariteit van
de kalibratielijn, selectiviteit, verloop van de kalibratielijn en
terugvinding van een toevoeging van fosfor aan grondwatermonsters en van
organisch gebonden fosfor. De resultaten van twee analyselijnen (P178 en
P213 nm), beiden gemeten met gebruik van de interne standaard elementen Ga
en In, werden met elkaar vergeleken. Uit deze experimenten kwam naar voren
dat de P213 nm lijn in combinatie met het interne standaard element Ga
superieur was in termen van gevoeligheid, aantoonbaarheidsgrens en
herhaalbaarheid.