Visser T ,
Vredenbregt MJ ,
Ginkel LA van ,
Schwillens PLWJ
14 p
in Dutch
1997
Toon Nederlands
English Abstract A pilot study has been carried out into the usefulness
of combined liquid chromatography and infrared spectrometry to confirm the
abuse of the anticoccidial drug Nicarbazin in samples of poultry feed. A
method has been developed based on the identification of the active
component of Nicarbazin, 4,4'-Dinitro-carbanilide (DNC). The principle is
concentration of the extracted sample on a trap column followed by
column-switching and back-flush gradient elution over a capillary analytical
column. IR-detection is carried out after elimination of the mobile phase
and deposition of the eluate on an IR transparant substrate by a spray-jet
interface. The limit of identification of DNC in samples is estimated as
40-60 ng injected, depending on the matrix. This corresponds with
approximately 1 mg/kg (1 ppm) in the original material.
Rapport in het kort
Orienterend onderzoek is uitgevoerd naar de
bruikbaarheid van vloeistofchromatografie met infraroodspectrometrische
detectie voor de bevestigingsanalyse van het cocciodio-staticum Nicarbazine
in afmestvoer voor pluimvee. Een methode is ontwikkeld gebaseerd op
identificatie van de actieve component van Nicarbazine, het
4,4'-Dinitro-carbanilide (DNC). Het principe van de bepaling is
concentratie van het monster op een voorkolom gevolgd door kolomschakeling
en 'back-flush' gradient-elutie over een capillaire analytische kolom.
IR-detectie vindt plaats na eliminatie van de mobiele fase en depositie van
de gescheiden componenten op een IR transparant medium met behulp van een
vernevelingsinterface. De identificatiegrens van DNC in monsters wordt
geschat op 40 tot 60 ng geinjecteerd, afhankelijk van de matrix. Dit komt
overeen met circa 1 mg/kg ( 1 ppm) in het oorspronkelijke
materiaal.