Go to abstract

Samenvatting

Abstract niet beschikbaar

Abstract

This report deals with the operationalization of a Perkin Elmer Zeeman 4100 ZL instrument for the determination of trace elements in rainwater. Choices were made for the settings of the various instrumental parameters. Performance characteristics were established and compared with the specifications supplied. The instrument met the suppliers specifications with respect to sensitivity, detection limits and precision. A comparative investigation with the formerly used equipment showed no significant or relevant differences in the results.

Resterend

Grootte
0MB